Serum protein electrophoresis by CZE 2000 clinical capillary electrophoresis system

Clin Chem. 1998 Apr;44(4):749-59.

Abstract

We compared the automated Paragon 2000 clinical capillary zone electrophoresis (CZE) system with two manual methods, agarose electrophoresis (AGE) and cellulose acetate electrophoresis (CAE). Reference intervals in healthy adults were determined for each method. When compared with AGE and CAE, CZE gave substantially higher reference values for the alpha1-globulin fraction. With CZE, within-run precision for fraction quantitation was between 0.5% (albumin) and 4.1% (alpha1-globulin). Total precision was between 0.8% (albumin) and 5.3% (beta-globulin). Data obtained from CZE showed poor linear correlation with results obtained by AGE but good linear correlation with data from CAE. Analysis of serum from patients with inter alia inflammation, nephrotic syndrome, or polyclonal gammopathy showed that clinical information obtained by CZE is comparable with information obtained by AGE and CAE. We conclude that CZE offers a clinically reliable alternative to AGE and CAE and has the advantages of automation, higher precision, and faster turnaround time.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Acute-Phase Reaction / blood
  • Adolescent
  • Adult
  • Aged
  • Bilirubin / analysis
  • Blood Protein Electrophoresis
  • Blood Proteins / analysis*
  • Diabetes Mellitus / blood
  • Electrophoresis, Agar Gel
  • Electrophoresis, Capillary / instrumentation
  • Electrophoresis, Cellulose Acetate
  • Female
  • Hemolysis
  • Humans
  • Lipids / analysis
  • Liver Diseases / blood
  • Male
  • Middle Aged
  • Nephrotic Syndrome / blood
  • Paraproteinemias / blood
  • Pregnancy
  • Reference Values
  • Time Factors

Substances

  • Blood Proteins
  • Lipids
  • Bilirubin