In vitro evaluation of three-dimensional ultrasonography based on magnetic scanhead tracking

Ultrasound Med Biol. 1998 Oct;24(8):1161-7. doi: 10.1016/s0301-5629(98)00098-2.

Abstract

The objective of this study was to evaluate the accuracy and precision of a magnetic position sensor system for acquisition of three-dimensional (3D) ultrasound images in volume estimation of phantoms in vitro. Installation of either 0.9% solution of saline at 37 degrees C or distilled water at 20 degrees C to a condom was performed. Scanning was performed either by a continuous or stepwise acquisition. This 3D ultrasound system demonstrated good correlation (r = 0.99-1.0, n = 8) between estimated (EV) and true volumes (TV). The errors were in the range 1.3%+/-0.3% (SEM) to 1.9%+/-0.6%, independent of sound velocity. Scanning through a porcine abdominal wall positioned at the fluid surface yielded a systematic underestimation of the volume: mean (EV - TV) = -7.2+/-0.8 ml. Eight repeated scans of the same volume yielded a coefficient of variation of 1.1%. Interobserver error of the tracing procedure was 2.6%+/-0.9%. This 3D ultrasound system gave high accuracy and precision in volume estimation in vitro, and yielded low interobserver error. A change in ultrasound velocity of approximately 60 m/s did not influence the accuracy significantly. Scanning through an abdominal wall underestimated volumes slightly.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Abdomen / diagnostic imaging*
  • Animals
  • Image Processing, Computer-Assisted / instrumentation*
  • Magnetics*
  • Observer Variation
  • Phantoms, Imaging*
  • Reproducibility of Results
  • Swine
  • Ultrasonography