Visualization of fluid turbulence and acoustic cavitation during phacoemulsification

J Cataract Refract Surg. 2005 Feb;31(2):406-11. doi: 10.1016/j.jcrs.2004.04.042.

Abstract

Purpose: To describe a technique for visualizing fluid turbulence and cavitational energy created by ultrasonic phaco tips.

Setting: University Eye Clinic of Trieste, Trieste, Italy.

Methods: Generation of cavitational energy by the phaco tip was visualized using an optical test bench comprising several components. The technique uses a telescope system to expand a laser light source into a coherent, collimated beam of light with a diameter of approximately 50.0 mm. The expanded laser beam shines on the test tube containing the tip activated in a medium of water or ophthalmic viscosurgical device (OVD). Two precision optical collimators complete the optical test bench and form the system used to focus data onto a charge-coupled device television camera connected to a recorder.

Results: Images of irrigation, irrigation combined with aspiration, irrigation/aspiration, and phacosonication were obtained with the tip immersed in a tube containing water or OVD.

Conclusions: Optical image processing enabled acoustic cavitation to be visualized during phacosonication. The system is a possible means of evaluating a single phaco apparatus power setting and comparing phaco machines and techniques.

MeSH terms

  • Acetates / chemistry*
  • Acoustics*
  • Drainage
  • Drug Combinations
  • Hyaluronic Acid / chemistry*
  • Minerals / chemistry*
  • Phacoemulsification / instrumentation*
  • Pressure*
  • Sodium Chloride / chemistry*
  • Therapeutic Irrigation
  • Vibration
  • Video Recording