Daily orthogonal kilovoltage imaging using a gantry-mounted on-board imaging system results in a reduction in radiation therapy delivery errors

Int J Radiat Oncol Biol Phys. 2012 Nov 1;84(3):596-601. doi: 10.1016/j.ijrobp.2012.01.033. Epub 2012 Apr 13.

Abstract

Purpose: To determine whether the use of routine image guided radiation therapy (IGRT) using pretreatment on-board imaging (OBI) with orthogonal kilovoltage X-rays reduces treatment delivery errors.

Methods and materials: A retrospective review of documented treatment delivery errors from 2003 to 2009 was performed. Following implementation of IGRT in 2007, patients received daily OBI with orthogonal kV X-rays prior to treatment. The frequency of errors in the pre- and post-IGRT time frames was compared. Treatment errors (TEs) were classified as IGRT-preventable or non-IGRT-preventable.

Results: A total of 71,260 treatment fractions were delivered to 2764 patients. A total of 135 (0.19%) TEs occurred in 39 (1.4%) patients (3.2% in 2003, 1.1% in 2004, 2.5% in 2005, 2% in 2006, 0.86% in 2007, 0.24% in 2008, and 0.22% in 2009). In 2007, the TE rate decreased by >50% and has remained low (P = .00007, compared to before 2007). Errors were classified as being potentially preventable with IGRT (e.g., incorrect site, patient, or isocenter) vs. not. No patients had any IGRT-preventable TEs from 2007 to 2009, whereas there were 9 from 2003 to 2006 (1 in 2003, 2 in 2004, 2 in 2005, and 4 in 2006; P = .0058) before the implementation of IGRT.

Conclusions: IGRT implementation has a patient safety benefit with a significant reduction in treatment delivery errors. As such, we recommend the use of IGRT in routine practice to complement existing quality assurance measures.

MeSH terms

  • Equipment Design
  • Humans
  • Medical Errors / classification
  • Medical Errors / prevention & control*
  • Medical Errors / statistics & numerical data
  • Particle Accelerators
  • Patient Identification Systems
  • Patient Safety
  • Quality Improvement*
  • Radiation Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy Setup Errors / prevention & control
  • Radiotherapy, Image-Guided / instrumentation
  • Radiotherapy, Image-Guided / methods*
  • Retrospective Studies