Comparison of Three Handheld Fundus Cameras for Assessment of the Vertical Cup-To-Disk Ratio

Ophthalmic Epidemiol. 2023 Sep 28:1-4. doi: 10.1080/09286586.2023.2260877. Online ahead of print.

Abstract

Purpose To compare the quality of optic nerve photographs from three different handheld fundus cameras and to assess the reproducibility and agreement of vertical cup-to-disk ratio (VCDR) measurements from each camera. Methods Adult patients from a comprehensive ophthalmology clinic and an intravitreous injection clinic in northern Thailand were recruited for this cross-sectional study. Each participant had optic nerve photography performed with each of 3 handheld cameras: the Volk iNview, Volk Pictor Plus, and Peek Retina. Images were graded for VCDR in a masked fashion by two photo-graders and images with > 0.2 discrepancy in VCDR were assessed by a third photo-grader. Results A total of 355 eyes underwent imaging with three different handheld fundus cameras. Optic nerve images were judged ungradable in 130 (37%) eyes imaged with Peek Retina, compared to 36 (10%) and 55 (15%) eyes imaged with the iNview and Pictor Plus, respectively. For 193 eyes with gradable images from all 3 cameras, inter-rater reliability for VCDR measurements was poor or moderate for each of the cameras, with intraclass correlation coefficients ranging from 0.41 to 0.52. A VCDR ≥ 0.6 was found in 6 eyes on iNview images, 9 eyes on Pictor Plus images, and 3 eyes on Peek images, with poor agreement between cameras (e.g., no eyes graded as VCDR ≥ 0.6 on images from both the iNview and Pictor Plus). Conclusions Inter-rater reliability of VCDR grades from 3 handheld cameras was poor. Cameras did not agree on which eyes had large VCDRs.

Keywords: Glaucoma; diagnostic techniques; ophthalmological; photography; reliability and validity.