Introduction: Accurate grading of pancreatic neuroendocrine tumors (PanNETs) relies on the assessment of Ki-67 immunohistochemistry (IHC). While digital imaging analysis (DIA) has been employed for Ki-67 IHC assessment in surgical specimens, its applicability to cytologic specimens remains underexplored. This study aimed to evaluate an automated DIA for assessing Ki-67 IHC on PanNET cell blocks.
Materials and methods: The study included 61 consecutive PanNETs and 5 pancreatic neuroendocrine carcinomas. Ki-67 IHC slides from cell blocks were digitally scanned into whole slide images using Philips IntelliSite Scanners and analyzed in batches using the Visiopharm Ki-67 App in a digital workflow. Ki-67 scores obtained through DIA were compared to pathologists' manual scores.
Results: The Pearson correlation coefficient of the percentage of Ki-67-stained nuclei between DIA reads and the originally reported reads was 0.9681. Concordance between DIA Ki-67 grades and pathologists' Ki-67 grades was observed in 92.4% (61/66) of cases with the calculated Cohen's Kappa coefficient of 0.862 (almost perfect agreement). Discordance between DIA and pathologists' consensus reads occurred in 5 PanNET cases which were upgraded from G1 to G2 by DIA due to contaminated Ki-67-stained inflammatory cells.
Conclusions: DIA demonstrated excellent concordance with pathologists' assessments, with only minor grading discrepancies. However, the essential role of pathologists in confirming results is emphasized to enhance overall accuracy.