Histological analysis of laser-enabled tissue coring with a novel 2910 nm erbium-doped fluoride glass fiber laser

Lasers Surg Med. 2024 May 8. doi: 10.1002/lsm.23792. Online ahead of print.

Abstract

Objectives: There remains an unmet need for a laser-enabled tissue coring device that can effectively improve face and neck skin laxity and rhytides. We investigate a novel 2910 nm erbium-doped fluoride glass fiber laser (2910 nm fiber laser) (UltraClear; Acclaro Medical) for laser-coring of submental tissue.

Methods: Five subjects, Glogau scale III-IV, were treated with a single pulse of the laser-coring mode of the 2910 nm fiber laser in the submentum. A 4 mm punch biopsy was immediately performed. Biopsy specimens were sectioned and stained with hematoxylin and eosin and placed on glass slides. All sections were reviewed, and sections containing the center of the transected core were analyzed for depth and diameter of the ablative microchannel and width of the surrounding zone of coagulation.

Results: A total of 15 intact micro-cores were analyzed. Histological analysis revealed an average ± standard deviation microchannel diameter of 242.5 ± 65.2 µm, an average ablative depth of 980 ± 318.8 µm, and an average zone of coagulation of 104 ± 32 µm.

Conclusions: Laser-enabled tissue coring with a novel 2910 nm fiber laser can safely achieve a wider microchannel diameter with ablative depth extending to the mid and deep dermis, which has the potential for collagen contraction and tissue tightening. Laser-coring to this ablation diameter and depth and with the surrounding zone of coagulation was found to be safe without adverse effects of post-inflammatory erythema or scarring in our study.

Keywords: 2910 nm; ablative fractional laser; laser coring; microcoring; tissue coring.