VISTECH contrast sensitivity testing in primary open angle glaucoma

Indian J Ophthalmol. 1997 Jun;45(2):99-103.

Abstract

Contrast sensitivity has been recommended as a screening and diagnostic test in primary open angle glaucoma (POAG). We tested contrast sensitivity (CS) using Vistech charts in 184 eyes of 95 patients. Three groups were examined--established primary open angle glaucoma, glaucoma suspects and age matched controls. The distribution of contrast sensitivities amongst the three groups were similar. The median contrast sensitivity of glaucoma suspects and controls were well within normal limits while that of the POAG group fell along the lower limit of normal. In all three groups the younger subjects scored better than the older, indicating a depression of contrast sensitivity with increasing age. Even if depression of any one spatial frequency was considered abnormal, the test yielded a sensitivity of 55.4% and specificity of 69.5%. Similarly contrast sensitivity testing was found to be of little use in detecting field defects a maximum sensitivity of 47.3% and specificity of 73.3%. Vistech contrast sensitivity testing is not a useful test in POAG screening or diagnosis.

MeSH terms

  • Adult
  • Aged
  • Contrast Sensitivity / physiology*
  • Glaucoma, Open-Angle / diagnosis
  • Glaucoma, Open-Angle / physiopathology*
  • Humans
  • Intraocular Pressure
  • Middle Aged
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Vision Disorders / diagnosis
  • Vision Disorders / physiopathology
  • Visual Fields