Reproducibility problems with the Abbott laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae

J Clin Microbiol. 2000 Jun;38(6):2416-8. doi: 10.1128/JCM.38.6.2416-2418.2000.

Abstract

This study demonstrates that significant reproducibility problems can occur during routine use of the Abbott Laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae. These problems can go undetected by the quality control procedures outlined in the manufacturer's package insert. We outline here procedures for detecting and preventing contamination and reproducibility problems.

Publication types

  • Clinical Trial

MeSH terms

  • Chlamydia Infections / diagnosis*
  • Chlamydia trachomatis / isolation & purification*
  • Gonorrhea / diagnosis*
  • Humans
  • Nucleic Acid Amplification Techniques*
  • Reagent Kits, Diagnostic*
  • Reproducibility of Results

Substances

  • Reagent Kits, Diagnostic