[Application of derivative technique in the signal processing of gas chromatography]

Se Pu. 1999 Mar;17(2):175-7.
[Article in Chinese]

Abstract

An application of derivative technique in the signal processing of gas chromatography(GC) is described. TCD's sensitivity of the derivative GC was 270 kV.L/kg, and its susceptibility was 0.3 microgram/L In comparison with the conventional GC, there was a 7-40 fold increase in sensitivity and a 24-40 fold decrease in susceptibility. The noise of the derivative GC was equal to or smaller than the conventional one. The minimum resolution for a pair of substances most difficulty to be separated in conventional GC, which was separated completely in the derivative GC, was 2.8. The linear range of n-butanol was 10(2) and the peak area correlation coefficient of the derivative GC was 0.9991.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • 1-Butanol / analysis
  • Chromatography, Gas / methods*
  • Sensitivity and Specificity

Substances

  • 1-Butanol