Quantitative detection of HIV-1 RNA using NucliSens EasyQ HIV-1 assay

J Virol Methods. 2005 Oct;129(1):40-6. doi: 10.1016/j.jviromet.2005.04.017.

Abstract

HIV-1 RNA viral load has become the major biological marker for disease prognosis and outcome of antiretroviral therapy in the treatment of HIV-infected individuals. The aim of this study was to compare the performance of the new CE marked NucliSens EasyQ HIV-1 assay with NucliSens HIV-1 QT assay (reference method). NucliSens EasyQ HIV-1 (EasyQ) couples nucleic acid sequence-based amplification (NASBA) with real-time detection using molecular beacons utilizing the NucliSens EasyQ analyzer. NASBA is a sensitive, isothermal, transcription-based amplification system designed specifically for the detection of RNA targets. There was significant correlation (r = 0.878, P < 0.0001) between the two different assays in the analysis of clinical samples and the frequency of concordant results (log difference <0.5) was 74%. The two assays detected HIV-1 RNA in 81 specimens, and neither detected (below the lower detection limit, 400 copies/ml for NucliSens HIV-1 QT and 500IU/ml for EasyQ) HIV-1 RNA in 12 specimens. Three clinical specimens had detectable HIV-1 RNA using the EasyQ only, and two specimens had detectable HIV-1 RNA using the NucliSens HIV-1 QT only. The EasyQ procedure can analyze 48 clinical samples within 6h. The coefficient of variation of EasyQ ranged from 3.0 to 9.5% (3% at 4.9 log; 5.7% at 3.7 log; 9.5% at 2.7 log). The new assay is shown to be a rapid, convenient, and reliable procedure for HIV-1 RNA viral load monitoring.

Publication types

  • Comparative Study

MeSH terms

  • HIV Infections / blood*
  • HIV Infections / virology
  • HIV-1 / genetics
  • HIV-1 / isolation & purification*
  • Humans
  • Microbiological Techniques
  • RNA, Viral / blood*
  • Reagent Kits, Diagnostic*
  • Sensitivity and Specificity
  • Viral Load

Substances

  • RNA, Viral
  • Reagent Kits, Diagnostic