Long-term optional retrievability of a new inferior vena cava filter in an ovine model

J Vasc Interv Radiol. 2005 Nov;16(11):1505-9. doi: 10.1097/01.RVI.0000180364.35781.48.

Abstract

Long-term retrievability of a new optional retrieval inferior vena cava (IVC) filter composed of a modified square stent and a modified Günther Tulip filter was tested in eight sheep. Eleven filters were placed into the IVC and eight were successfully retrieved 3-5 months after implantation. Incorporation of the filter struts into the IVC wall prevented its retrieval in three sheep at 3, 4, and 5 months after placement.

MeSH terms

  • Animals
  • Blood Vessel Prosthesis Implantation
  • Device Removal*
  • Disease Models, Animal
  • Female
  • Follow-Up Studies
  • Phlebography
  • Prosthesis Design / classification
  • Sheep
  • Stents / classification
  • Time Factors
  • Vascular Patency
  • Vena Cava Filters*
  • Vena Cava, Inferior / diagnostic imaging
  • Vena Cava, Inferior / pathology
  • Vena Cava, Inferior / surgery*