Large-format imaging plate and weissenberg camera for accurate protein crystallographic data collection using synchrotron radiation

J Synchrotron Radiat. 1997 May 1;4(Pt 3):136-46. doi: 10.1107/S0909049597003683.

Abstract

Off-line and on-line protein data-collection systems using an imaging plate as a detector are described and their components reported. The off-line scanner IPR4080 was developed for a large-format imaging plate ;BASIII' of dimensions 400 x 400 mm and 400 x 800 mm. The characteristics of this scanner are a dynamic range of 10(5) photons pixel(-1), low background noise and high sensitivity. A means of reducing electronic noise and a method for finding the origin of the noise are discussed in detail. A dedicated screenless Weissenberg camera matching IPR4080 with synchrotron radiation was developed and installed on beamline BL6B at the Photon Factory. This camera can attach one or two sheets of 400 x 800 mm large-format imaging plate inside the film cassette by evacuation. The positional reproducibility of the imaging plate on the cassette is so good that the data can be processed by batch job. Data of 93% completeness up to 1.6 A resolution were collected on a single axis rotation and the value of R(merge) becomes 4% from a tetragonal lysozyme crystal using a set of two imaging-plate sheets. Comparing two types of imaging plates, the signal-to-noise ratio of the ST-VIP-type imaging plate is 25% better than that of the BASIII-type imaging plate for protein data collection using 1.0 and 0.7 A X-rays. A new on-line protein data-collection system with imaging plates is specially designed to use synchrotron radiation X-rays at maximum efficiency.