Rotary powered device for bone marrow aspiration and biopsy yields excellent specimens quickly and efficiently

J Clin Pathol. 2010 Jun;63(6):562-5. doi: 10.1136/jcp.2009.072843. Epub 2010 Apr 19.

Abstract

Recently, a new FDA-cleared battery powered bone marrow biopsy system was developed to allow operators access to the bone marrow space quickly and efficiently. A pre-clinical evaluation of the device (OnControl, Vidacare Corporation, San Antonio, TX, USA) on anesthetized pigs was conducted, in addition to a clinical evaluation in hematology clinic patients requiring a bone marrow biopsy. Twenty-six samples were collected from the swine model. No cellular artifact or thermal damage was reported in any of the samples obtained. For the clinical evaluation of the device, 16 patients were recruited. Mean time from needle contact with skin to needle removal was 38.5 +/- 13.94 seconds. No complications were reported. In this study, the manual and powered samples were equivalent in specimen quality. In the patients evaluated, the device was safe, easy to use and the mean procedural time was significantly faster than previously reported with a manual technique.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Biopsy, Needle / instrumentation*
  • Biopsy, Needle / methods
  • Bone Marrow / pathology*
  • Bone Marrow Examination / instrumentation*
  • Bone Marrow Examination / methods
  • Equipment Design
  • Humans
  • Models, Animal
  • Sus scrofa