Tacrolimus-eluting carbon-coated stents versus sirolimus-eluting stents for prevention of symptom-driven clinical end points

Clin Res Cardiol. 2010 Oct;99(10):645-50. doi: 10.1007/s00392-010-0165-3. Epub 2010 Apr 20.

Abstract

Background: Coating of stents has been shown to minimize the interactions between platelets, stent surface and vascular response following stent implantation. The aim of our study was to compare the tacrolimus-eluting carbon-coated JANUS(®) stent with sirolimus-eluting CYPHER(®) stent for the prevention of symptom-driven clinical end points in a real world clinical setting.

Methods: This prospective registry with a follow-up period of 24 months was conducted in 90 consecutive patients undergoing coronary artery stenting receiving CYPHER(®) (n = 48) or JANUS(®) (n = 42) stents. The primary end point was a composite of death from cardiovascular causes, nonfatal myocardial infarction and target vessel revascularisation, and the secondary end point was clinically driven in-stent restenosis.

Results: The primary combined endpoint occurred in 38% of patients (n = 16) in the JANUS(®) group compared to 10% (n = 5) in the CYPHER(®) group. The relative risk increase of the composite end point was therefore 63% higher in patients receiving JANUS(®) stents compared to the CYPHER(®) stents (crude HR = 1.63, 95% CI = 1.17-2.28, p = 0.004; adjusted HR = 1.79, CI = 1.26-2.55, p = 0.001). Interestingly, 75% of events in the JANUS(®) group occurred during the first 6 months after stent implantation. Similarly, the rate of clinically driven in-stent restenosis was higher in patients receiving JANUS(®) stent (n = 10, 2%) compared to the CYPHER(®) stent (n = 2, 4%). Concordantly, the relative risk for clinically driven in-stent restenosis was 81% higher in the JANUS(®) group compared to the CYPHER(®) group (crude HR = 1.81, 95% CI = 1.08-3.02, p = 0.02; adjusted HR = 2.24, CI = 1.26-3.96, p = 0.006).

Conclusion: The use of tacrolimus-eluting carbon coated JANUS(®) stent was associated with worse clinical outcome compared to the sirolimus-eluting CYPHER(®) stent in clinical routine use.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary*
  • Carbon
  • Cardiovascular Diseases / prevention & control*
  • Coronary Disease / therapy*
  • Drug-Eluting Stents / adverse effects*
  • Female
  • Humans
  • Immunosuppressive Agents / administration & dosage*
  • Male
  • Middle Aged
  • Registries
  • Sirolimus / administration & dosage
  • Tacrolimus / administration & dosage
  • Treatment Outcome

Substances

  • Immunosuppressive Agents
  • Carbon
  • Sirolimus