Successful retrieval of an impacted mechanical lithotripsy basket: a case report

Dig Endosc. 2010 Jul:22 Suppl 1:S111-3. doi: 10.1111/j.1443-1661.2010.00963.x.

Abstract

We report the successful retrieval of an impacted mechanical lithotripsy basket. In a patient with two large common bile duct stones, the basket with the entrapped stone was impacted within the mid-common bile duct. We then attempted to use another mechanical lithotripter; however the central wire of the basket fractured at the handle portion. Grasping a few wires of the impacted basket with rat-tooth forceps allowed the wires of the basket to slip away from the stone. The present report describes the safe and effective use of rat-tooth forceps in the management of an impacted lithotripter basket.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cholangiopancreatography, Endoscopic Retrograde
  • Device Removal
  • Equipment Failure
  • Female
  • Fluoroscopy
  • Gallstones / diagnosis
  • Gallstones / therapy*
  • Humans
  • Lithotripsy / adverse effects*
  • Lithotripsy / instrumentation
  • Reoperation / methods*