Randomized trial comparing the i-gel™ and Magill tracheal tube with the single-use ILMA™ and ILMA™ tracheal tube for fibreoptic-guided intubation in anaesthetized patients with a predicted difficult airway

Br J Anaesth. 2011 Aug;107(2):251-7. doi: 10.1093/bja/aer103. Epub 2011 Jun 7.

Abstract

Background: The i-gel™ is a single-use supraglottic airway device (SAD) that allows fibreoptic-guided tracheal intubation through the device. Until now, no prospective data for this procedure are available. Therefore, in a prospective randomized controlled trial, we evaluated fibreoptic-guided tracheal intubation with a standard Rüsch™ PVC tracheal tube (TT) through the i-gel™ compared with the single-use ILMA™ (sILMA™) TT through the sILMA™ in patients with a predicted difficult airway.

Methods: With ethics committee approval and written informed consent, 160 patients were randomly assigned to either SAD. After placement of the SAD, a fibreoptic bronchoscope was introduced into the trachea as a railroad for the TT. Primary outcome variable was the first-attempt fibreoptic-guided intubation success rate. Secondary variables included time for insertion and intubation, airway leak pressures, fibreoptic view, and adverse events. Data are presented as mean (sd) or percentages. A P-value of < 0.05 was considered statistically significant.

Results: Fibreoptic-guided intubation was successful at the first attempt in 76 patients (96%) using the i-gel™ and in 71 patients (90%) using the sILMA™ (P=0.21). Most of the failed intubations were rescued by conventional laryngoscopy. Airway leak pressure was higher for the sILMA™. There were no problems during removal of either type of SAD.

Conclusions: Fibreopic-guided tracheal intubation through the i-gel™ using a standard Rüsch™ Magill TT is successful and an alternative to the sILMA™ with the sILMA™ TT.

Publication types

  • Comparative Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Anesthesia, General / methods
  • Device Removal
  • Disposable Equipment
  • Equipment Design
  • Female
  • Fiber Optic Technology / methods*
  • Humans
  • Intubation, Intratracheal / adverse effects
  • Intubation, Intratracheal / instrumentation*
  • Laryngeal Masks / adverse effects
  • Male
  • Middle Aged
  • Pharyngitis / etiology
  • Postoperative Complications
  • Prospective Studies
  • Single-Blind Method
  • Young Adult