Troponin-specific autoantibody interference in different cardiac troponin I assay configurations

Clin Chem. 2012 Jun;58(6):1040-8. doi: 10.1373/clinchem.2011.179226. Epub 2012 Apr 9.

Abstract

Background: Autoantibodies to cardiac troponins (cTnAAb) can interfere with the measurement of cardiac troponin I (cTnI) by immunoassays. The aim of this study was to explore the degree of cTnAAb interference in different cTnI assay configurations.

Methods: Ternary troponin complex was added into samples (serum or plasma, n = 132, 68% cTnAAb positive) from individuals without known cardiac conditions. The recovery of cTnI was then measured with 6 investigational cTnI assays (2, 3, or 4 antibodies per assay). Three of these assays were then selected for further comparison by use of samples (plasma, n = 210, 33% cTnAAb positive) from non-ST-elevation acute coronary syndrome patients in the FRISC-II (FRagmin/Fast Revascularisation during InStability in Coronary artery disease) cohort. Finally, these results were compared to those obtained with 3 commercial cTnI assays.

Results: Analytical recoveries varied widely among the 6 investigational assays. Notably the low recoveries (median 9%) of the midfragment-targeting reference assay were normalized (median 103%) with the use of the 4-antibody assay construct (3 capture, 1 tracer antibody) with only 1 antibody against a midfragment epitope. Reduced analytical recoveries correlated closely with measured autoantibody amounts. cTnI concentrations from cTnAAb-positive patient samples determined with 3 investigational assays confirmed the reduced concentrations expected from the low analytical recoveries. The results from the commercial cTnI assays with antibody selections representative for contemporary assay constructs revealed a similar underestimation (up to 20-fold) of cTnI in cTnAAb-positive samples.

Conclusions: A novel cTnI assay deviating from the conventional IFCC-recommended midfragment approach substantially improves cTnI detection in samples containing cTnAAbs.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acute Coronary Syndrome / blood
  • Autoantibodies / blood*
  • Epitopes
  • False Negative Reactions
  • False Positive Reactions
  • Humans
  • Immunoassay
  • Multicenter Studies as Topic
  • Prospective Studies
  • Sensitivity and Specificity
  • Troponin I / blood
  • Troponin I / immunology*

Substances

  • Autoantibodies
  • Epitopes
  • Troponin I