Results with the Talent thoracic stent graft in the VALOR trial

J Vasc Surg. 2012 Nov;56(5):1214-21.e1. doi: 10.1016/j.jvs.2012.04.071. Epub 2012 Aug 25.

Abstract

Objective: We report the 5-year outcomes of thoracic endovascular aneurysm repair (TEVAR) using the Medtronic Vascular Talent Thoracic Stent Graft System (Medtronic Vascular, Santa Rosa, Calif) in patients considered low or moderate risk for open surgical repair.

Methods: The Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial was a prospective, nonrandomized, multicenter, pivotal study conducted at 38 U.S. sites. Between December 2003 and June 2005, VALOR enrolled 195 patients who were low or moderate risk (0, 1, and 2) per the modified Society for Vascular Surgery and American Association for Vascular Surgery criteria. The patients had fusiform thoracic aortic aneurysms (TAAs) and/or focal saccular TAAs/penetrating atherosclerotic ulcers. Standard follow-up interval examinations were conducted at 1 month, 6 months, 1 year, and annually thereafter.

Results: Over the 5-year follow-up, 76 deaths occurred (43.9%). Freedom from all-cause mortality was 83.9% at 1 year and 58.5% at 5 years. Most deaths were due to cardiac, pulmonary or cancer-related causes. Freedom from aneurysm-related mortality (ARM) was 96.9% at 1 year and 96.1% at 5 years. There was only 1 case of ARM after the first year of follow-up. Over the 5-year follow-up period, four patients were converted to open surgery and four patients experienced aneurysm rupture. The 5-year freedom from aneurysm rupture was 97.1% and the 5-year freedom from conversion to surgery was 97.1%. The incidence of stent graft migration (>10 mm) was ≤ 1.8% in each year of follow-up. The rate of type I endoleak was 4.6% at 1 month, 6.3% from 1 month to 1 year, and 3.8% during year 5. The rate of type III endoleak was 1.3% at 1 month, 1.9% from 1 month to 1 year, and 1.9% during year 5. Through 5 years, 28 patients (14.4%) underwent 31 additional endovascular procedures on the original target lesion. The 5-year freedom from secondary endovascular procedures was 81.5%.

Conclusions: Through 5-year follow-up in patients who were candidates for open surgical repair, TEVAR using the Talent Thoracic Stent Graft System has demonstrated sustained protection from ARM, aneurysm rupture, and conversion to surgery, and durable stent graft performance. Close patient follow-up remains essential after TEVAR.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aortic Aneurysm, Thoracic / surgery*
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / methods*
  • Follow-Up Studies
  • Humans
  • Postoperative Complications / epidemiology
  • Postoperative Complications / etiology
  • Prospective Studies
  • Stents*