Dose calibrator manufacturer-dependent bias in assays of ¹²³I

Appl Radiat Isot. 2014 Aug:90:79-83. doi: 10.1016/j.apradiso.2014.03.018. Epub 2014 Mar 26.

Abstract

Calibration factors for commercial ionization chambers (i.e. dose calibrators) were determined for a solution of (123)I; the activity was based on the 1976 NBS standard. A link between the NIST standard and the International Reference System (SIR) was established. The two major U.S. dose calibrator manufacturers recommend oppositely biased calibration factors, giving a spread of 11.3% in measured activities. With modern quantitative imaging techniques capable of ≤10% accuracy, this bias for a SPECT nuclide is highly significant.

Keywords: Dose calibrator; I-123; Ionization chamber; Quantitative imaging; Secondary standard.