Platelet activation is less enhanced in the new balloon expandable Edwards Sapien 3 valve compared to its predecessor model (Edwards Sapien XT)

Thromb Haemost. 2016 Jan;115(1):109-16. doi: 10.1160/TH15-03-0267. Epub 2015 Aug 27.

Abstract

Stroke and thromboembolic events after transfemoral aortic valve replacement (TAVR) continue to be a problem. The aim of our study was to compare platelet aggregation (Agg) and platelet activation (PA) observed with two different catheter valves, the ESV-XT and the newer ESV-3 valve in patients (pts) undergoing TAVR on dual antiplatelet therapy (DAPT). A total of 174 patients with severe aortic stenosis and high surgical risk successfully underwent TAVR (60 ESV-XT; 114 ESV-3). Platelet Agg and PA (CD62P expression) were evaluated before and the following three days after TAVR under DAPT. Platelet Agg was inhibited to the same extent in both valve types and there was no significant difference in platelet drop between both valve types between day 0 and day 3 [ESV-XT vs ESV-3: median (25th-75th percentile): platelet count (x1000): 55 (42-74) vs 61(42-93), p=0.280]. However, there was an enhanced CD62P expression directly after TAVR with the ESV-XT compared to the ESV-3 [CD62P (MIF): 7.4 (6.8-8.6) vs 6.6 (6-7.9), p=0.014]. Surface expression of platelet CD62P was associated with the occurrence of residual aortic regurgitation (AR) and was significantly higher in patients with residual AR [CD62P (mild AR) vs CD 62P (no or trace AR): 7.9 (7.3-9.1) vs 7.1 (6.4-8.0), p < 0.001)]. PA was significantly enhanced in patients with the ESV-XT compared to the ESV-3 valve and was associated with the amount of residual AR which was significantly reduced by ESV-3. This may have implications for thromboembolic events following TAVR procedure.

Keywords: DAPT; Edwards Sapien 3 valve; Edwards Sapien XT; Platelet activation; TAVR; residual aortic regurgitation.

Publication types

  • Comparative Study
  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged, 80 and over
  • Aortic Valve Insufficiency / blood
  • Aortic Valve Insufficiency / diagnosis
  • Aortic Valve Insufficiency / therapy*
  • Aortic Valve Stenosis / blood
  • Aortic Valve Stenosis / diagnosis
  • Aortic Valve Stenosis / therapy*
  • Aortic Valve*
  • Balloon Valvuloplasty*
  • Biomarkers / blood
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Drug Therapy, Combination
  • Female
  • Germany
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • P-Selectin / blood
  • Platelet Activation* / drug effects
  • Platelet Aggregation
  • Platelet Aggregation Inhibitors / therapeutic use
  • Platelet Count
  • Platelet Function Tests
  • Prospective Studies
  • Prosthesis Design
  • Risk Factors
  • Severity of Illness Index
  • Time Factors
  • Treatment Outcome

Substances

  • Biomarkers
  • P-Selectin
  • Platelet Aggregation Inhibitors
  • SELP protein, human