The laboratory analysis provides accurate, but time consuming hemoglobin level estimation especially in the emergency setting. The reliability of time-sparing point of care devices (POCT) remains uncertain. We tested two POCT devices accuracy (HemoCue®201+ and Gem®Premier™3000) in routine emergency department workflow. Blood samples taken from patients admitted to the emergency department were analyzed for hemoglobin concentration using a laboratory reference Beckman Coulter LH 750 (HBLAB), the HemoCue (HBHC) and the Gem Premier 3000 (HBGEM). Pairwise comparison for each device and HbLAB was performed using correlation and the Bland-Altman methods. The reliability of transfusion decision was assessed using three-zone error grid. A total of 292 measurements were performed in 99 patients. Mean hemoglobin level were 115 ± 33, 110 ± 28 and 111 ± 30 g/l for HbHC, HbGEM and HbLAB respectively. A significant correlation was observed for both devices: HbHC versus HbLAB (r2 = 0.93, p < 0.001) and HBGEM versus HBLAB (r2 = 0.86, p < 0.001). The Bland-Altman method revealed bias of -3.7 g/l (limits of agreement -20.9 to 13.5) for HBHC and HBLAB and 2.5 g/l (-18.6 to 23.5) for HBGEM and HBLAB, which significantly differed between POCT devices (p < 0.001). Using the error grid methodology: 94 or 91 % of values (HbHC and HbGEM) fell in the zone of acceptable difference (A), whereas 0 and 1 % (HbHC and HbGEM) were unacceptable (zone C). The absolute accuracy of tested POCT devices was low though reaching a high level of correlation with laboratory measurement. The results of the Morey´s error grid were unfavorable for both POCT devices.
Keywords: Blood transfusion; Emergency medicine; Hemoglobin; Hemoglobinometry; Hemorrhage; Point of care systems.