Thin strut bare metal stents in patients with atrial fibrillation: Is there still a need for BMS?

Catheter Cardiovasc Interv. 2016 Sep;88(3):358-66. doi: 10.1002/ccd.26261. Epub 2015 Dec 9.

Abstract

Objectives: This observational study assessed the 9-month clinical outcomes in an « all comers » population with a focus on patients with atrial fibrillation (AF) after thin strut bare metal stenting.

Background: Drug eluting stent (DES) implantation is the treatment of choice for coronary artery disease (CAD) leaving only marginal indications for the use of bare metal stents (BMS). However, selected treatment populations with DES contraindications such as patients who cannot sustain 6-12 months of dual antiplatelet therapy (DAPT) remain candidates for BMS implantations.

Methods: Thin strut bare metal stenting in a priori defined subgroups were investigated in a non-randomized, international, multicenter «all comers» observational study. Primary endpoint was the 9-month TLR rate whereas secondary endpoints included the 9-month MACE and procedural success rates.

Results: A total of 783 patients of whom 98 patients had AF underwent BMS implantation. Patient age was 70.4 ± 12.8 years. Cardiovascular risk factors in the overall population were male gender (78.2%, 612/783), diabetes (25.2%, 197/783), hypertension (64.1%, 502/783), cardiogenic shock (4.9%, 38/783) and end stage renal disease (4.9%, 38/783). In-hospital MACE was 4.1% (30/783) in the overall population. The 9-month TLR rate was 4.5% (29/645) in the non-AF group and 3.3% (3/90) in the AF group (P = 0.613). At 9 months, the MACE rate in the AF-group and non-AF group was not significantly different either (10.7%, 69/645 vs. 6.7%, 6/90; P = 0.237). Accumulated stroke rates were 0.3% (2/645) in the non-AF subgroup at baseline and 1.1% (1/90) in the AF subgroup (P = 0.264).

Conclusion: Bare metal stenting in AF patients delivered acceptably low TLR and MACE rates while having the benefit of a significantly shorter DAPT duration in a DES dominated clinical practice. © 2015 Wiley Periodicals, Inc.

Keywords: MACE; TLR; all comers observational study; atrial fibrillation; bare metal stent.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Anticoagulants / administration & dosage
  • Asia
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / etiology*
  • Atrial Fibrillation / mortality
  • Atrial Fibrillation / therapy
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Drug Therapy, Combination
  • Europe
  • Female
  • Hemorrhage / chemically induced
  • Humans
  • Male
  • Metals*
  • Middle Aged
  • Myocardial Infarction / etiology
  • Percutaneous Coronary Intervention / adverse effects*
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Platelet Aggregation Inhibitors / administration & dosage
  • Prosthesis Design
  • Risk Factors
  • Stents*
  • Stroke / etiology
  • Time Factors
  • Treatment Outcome

Substances

  • Anticoagulants
  • Metals
  • Platelet Aggregation Inhibitors