RM Pressfit® cup: good preliminary results at 5 to 8 years follow-up for 189 patients

Hip Int. 2016 Jul 25;26(4):386-91. doi: 10.5301/hipint.5000359. Epub 2016 May 30.

Abstract

Aim: The RM Classic® cup shows very good results. The RM Pressfit® cup has an equatorial pressfit; it is a monoblock implant without metalback. The aim of this study was to evaluate retrospectively the survival of this implant, monitoring patients for at least 5 years in all consecutive cases operated in our centre.

Methods: Between February 2006 and December 2008, our department performed 189 consecutive nonselected primary total hip arthroplasties using the RM® Pressfit cup, and all of these were included in the study. The operative technique involved an anterolateral approach. The latest follow-up evaluation was performed at a minimum of 5 years after the index arthroplasty.

Results: The mean follow-up was 6.5 years (5-8 years). We had preoperative information for 186 of the 189 patients. 102 patients were seen in consultation (54.0%), 46 (24.3%) patients were contacted by phone, 37 (19.6%) patients died, and 4 patients (2.1%) were lost for unknown reasons. Of the 189 arthroplasties, 6 (3.2%) required revision; 4 (2.1%) for infection, 1 (0.5%) for significant haematoma with persistent deglobulisation and 1 (0.5%) for periprosthetic calcification during the 5 to 8 years following the operation. No hip was revised because of aseptic loosening. The mean annual wear rate based on the last follow-up at 5 years minimum was 0.065 mm per year. We observed no migration, 1 femoral loosening and no acetabular loosening.

Conclusions: Our study finds a good survival rate which is compatible with the 94% at 20 years with RM Classic®. These results are suggestive but must be confirmed with long-term studies.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Female
  • Follow-Up Studies
  • Hip Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / etiology
  • Osteoarthritis, Hip / surgery*
  • Prosthesis Design*
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome