Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood

Ann Lab Med. 2017 Mar;37(2):147-150. doi: 10.3343/alm.2017.37.2.147.

Abstract

There has been increasing interest in standardized and quantitative Epstein-Barr virus (EBV) DNA testing for the management of EBV disease. We evaluated the performance of the Real-Q EBV Quantification Kit (BioSewoom, Korea) in whole blood (WB). Nucleic acid extraction and real-time PCR were performed by using the MagNA Pure 96 (Roche Diagnostics, Germany) and 7500 Fast real-time PCR system (Applied Biosystems, USA), respectively. Assay sensitivity, linearity, and conversion factor were determined by using the World Health Organization international standard diluted in EBV-negative WB. We used 81 WB clinical specimens to compare performance of the Real-Q EBV Quantification Kit and artus EBV RG PCR Kit (Qiagen, Germany). The limit of detection (LOD) and limit of quantification (LOQ) for the Real-Q kit were 453 and 750 IU/mL, respectively. The conversion factor from EBV genomic copies to IU was 0.62. The linear range of the assay was from 750 to 10⁶ IU/mL. Viral load values measured with the Real-Q assay were on average 0.54 log₁₀ copies/mL higher than those measured with the artus assay. The Real-Q assay offered good analytical performance for EBV DNA quantification in WB.

Keywords: Epstein-Barr virus; Performance; Quantification; Real-time PCR; Whole blood.

Publication types

  • Evaluation Study

MeSH terms

  • DNA, Viral / blood*
  • DNA, Viral / metabolism
  • Epstein-Barr Virus Infections / diagnosis
  • Epstein-Barr Virus Infections / virology
  • Herpesvirus 4, Human / genetics*
  • Herpesvirus 4, Human / isolation & purification
  • Humans
  • Limit of Detection
  • Reagent Kits, Diagnostic
  • Real-Time Polymerase Chain Reaction

Substances

  • DNA, Viral
  • Reagent Kits, Diagnostic