Four-dimensional cone-beam computed tomography-guided radiotherapy for gastric lymphoma

Jpn J Radiol. 2018 Feb;36(2):159-163. doi: 10.1007/s11604-017-0698-8. Epub 2017 Oct 30.

Abstract

Purpose: We describe a treatment method with four-dimensional cone-beam computed tomography (4D-CBCT)-guided radiotherapy for gastric lymphoma.

Materials and methods: We performed image-guided radiotherapy (IGRT) with 15 fractions for a gastric mucosa-associated lymphoid tissue lymphoma patient, using 4D-CBCT. The stomach was delineated based on 4D-CT images. For image guidance, an automatic registration between planning CT and 4D-CBCT images was performed based on the bony anatomy (bone matching), followed by manual registration based on the stomach in 4D-CBCT images of all 10 phases (4D matching). We calculated the covering ratio (CR) with variable stomach-to-planning target volume (PTV) margins, based on the images of all phases [CR (%) = the number of covering phases/all 150 phases × 100].

Results: The patient underwent radiotherapy (RT) as scheduled, without any significant adverse effects. The appropriate PTV margins (CR ≥ 95%) were 25 mm (CR 99.3%) and 15 mm (CR 98.7%) for bone and 4D matching, respectively.

Conclusion: 4D matching using 4D-CBCT is appropriate for IGRT of gastric lymphomas.

Keywords: Four-dimensional cone-beam computed tomography; Gastric lymphoma; Image-guided radiotherapy; Internal target volume; Respiratory motion.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cone-Beam Computed Tomography / methods*
  • Female
  • Four-Dimensional Computed Tomography / methods*
  • Humans
  • Lymphoma, Non-Hodgkin / diagnostic imaging*
  • Lymphoma, Non-Hodgkin / radiotherapy*
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy, Image-Guided / methods*
  • Stomach / diagnostic imaging
  • Stomach Neoplasms / diagnostic imaging*
  • Stomach Neoplasms / radiotherapy*

Supplementary concepts

  • Familial primary gastric lymphoma