Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details?

Eur Radiol. 2018 Dec;28(12):5051-5059. doi: 10.1007/s00330-018-5496-x. Epub 2018 May 28.

Abstract

Objectives: To compare contrast-to-noise ratios (CNRs) and iodine discrimination thresholds on iodine maps derived from dual energy CT (DECT) and subtraction CT (SCT).

Methods: A contrast-detail phantom experiment was performed with 2 to 15 mm diameter tubes containing water or iodinated contrast concentrations ranging from 0.5 mg/mL to 20 mg/mL. DECT scans were acquired at 100 kVp and at 140 kVp+Sn filtration. SCT scans were acquired at 100 kVp. Iodine maps were created by material decomposition (DECT) or by subtraction of water scans from iodine scans (SCT). Matched exposure levels varied from 8 to 15 mGy. Iodine discrimination thresholds (Cr) and response times were determined by eight observers.

Results: The adjusted mean CNR was 1.9 times higher for SCT than for DECT. Exposure level had no effect on CNR. All observers discriminated all details ≥10 mm at 12 and 15 mGy. For sub-centimetre details, the lowest calculated Cr was ≤ 0.50 mg/mL for SCT and 0.64 mg/mL for DECT. The smallest detail was discriminated at ≥4.4 mg/mL with SCT and at ≥7.4 mg/mL with DECT. Response times were lower for SCT than DECT.

Conclusions: SCT results in higher CNR and reduced iodine discrimination thresholds compared to DECT for sub-centimetre details.

Key points: • Subtraction CT iodine maps exhibit higher CNR than dual-energy iodine maps • Lower iodine concentrations can be discriminated for sub-cm details with SCT • Response times are lower using SCT compared to dual-energy CT.

Keywords: Contrast Media; Iodine; Phantoms, Imaging; Subtraction technique; Tomography, X-Ray Computed.

Publication types

  • Comparative Study

MeSH terms

  • Contrast Media
  • Humans
  • Iodine
  • Phantoms, Imaging
  • Radiographic Image Interpretation, Computer-Assisted / methods
  • Radiography, Dual-Energy Scanned Projection / methods*
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Iodine