Simultaneous pace-ablate during CARTO-guided pulmonary vein isolation with a contact-force sensing radiofrequency ablation catheter

J Interv Card Electrophysiol. 2019 Mar;54(2):119-124. doi: 10.1007/s10840-018-0455-7. Epub 2018 Sep 27.

Abstract

Purpose: Elimination of pace-capture along pulmonary vein isolation (PVI) lesion sets reduces atrial fibrillation (AF) recurrence in catheter ablation of paroxysmal AF. Pacing from the RF ablation electrode during RF application is prevented within the CARTO electroanatomic mapping system (Biosense Webster, Inc.) due to theoretical safety considerations. We evaluated a method of pacing the distal ablation electrode during RF application in the CARTO system, thus avoiding repeated activation and inactivation of the pacing channel and facilitating immediate recognition of pace-capture loss. We investigated the safety, feasibility, and utility of simultaneous pace-ablate (SPA) during AF ablation with the CARTO-3 system and a contact-force sensing RF ablation catheter.

Methods: Safety of feasibility of SPA was evaluated in 250 patients undergoing first-time AF ablation. Frequency and regional distribution of pace-capture following PVI was evaluated in a cohort of 50 consecutive patients undergoing catheter ablation of paroxysmal AF.

Results: SPA was successfully performed in all 250 patients without adverse event. At least one pace-capture site was noted in 22 of 50 PAF patients (44%), and pace-capture following PVI was most common at anterior and superior left atrial sites. There were 2.0 ± 3.3 RF applications during pacing via the distal ablation electrode per patient, and all lesions sets were successfully rendered unexcitable.

Conclusions: Pace-capture along the completed PVI lesion set remains common despite utilization of contact-force sensing RF ablation catheters and automated lesion annotation. Simultaneous pace-ablate in AF ablation using the CARTO system may be safely used to render atrial lesion sets unexcitable.

Keywords: Atrial fibrillation; Catheter ablation; Contact-force sensing; Electroanatomic mapping; Pacing; Radiofrequency ablation.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / surgery*
  • Cardiac Catheters*
  • Cardiac Pacing, Artificial / methods
  • Cohort Studies
  • Combined Modality Therapy
  • Electrocardiography / methods
  • Equipment Design
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Middle Aged
  • Patient Safety / statistics & numerical data*
  • Prognosis
  • Pulmonary Veins / surgery*
  • Radiofrequency Ablation / methods*
  • Retrospective Studies
  • Risk Assessment
  • Surgery, Computer-Assisted / instrumentation
  • Treatment Outcome