A truly non-occlusive stent-graft moulding balloon for thoracic endovascular aortic repair

Interact Cardiovasc Thorac Surg. 2019 Sep 1;29(3):352-354. doi: 10.1093/icvts/ivz098.

Abstract

Endovascular treatment of thoracic aortic pathologies has recently progressed towards more proximal pathologies, including those of the aortic arch and ascending aorta where there is a higher risk for stent-graft migration during the deployment or the moulding procedure due to the beating and ejecting heart. Typical measures to prevent dislodgement of the balloon or the stent-graft during the procedure are rapid pacing or pharmacologically induced hypotension. We present a circular and fully non-occlusive stent-graft moulding balloon that does not require any reduction of cardiac output or hypotension during inflation, moulding and deflation of the balloon.

Keywords: TEVAR; balloon-moulding; endoleak; non-occlusive; stent-graft.

MeSH terms

  • Aorta, Thoracic / surgery
  • Aortic Aneurysm, Thoracic / surgery*
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis Implantation / methods
  • Blood Vessel Prosthesis*
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / methods
  • Equipment Design
  • Humans
  • Stents*
  • Time Factors
  • Treatment Outcome