Feasibility of mixed reality-based intraoperative three-dimensional image-guided navigation for atlanto-axial pedicle screw placement

Proc Inst Mech Eng H. 2019 Dec;233(12):1310-1317. doi: 10.1177/0954411919881255. Epub 2019 Oct 16.

Abstract

This study aimed to evaluate the safety and accuracy of mixed reality-based intraoperative three-dimensional navigated pedicle screws in three-dimensional printed model of fractured upper cervical spine. A total of 27 cervical model from patients of upper cervical spine fractures formed the study group. All the C1 and C2 pedicle screws were inserted under mixed reality-based intraoperative three-dimensional image-guided navigation system. The accuracy and safety of the pedicle screw placement were evaluated on the basis of postoperative computerized tomography scans. A total of 108 pedicle screws were properly inserted into the cervical three-dimensional models under mixed reality-based navigation, including 54 C1 pedicle screws and 54 C2 pedicle screws. Analysis of the dimensional parameters of each pedicle at C1/C2 level showed no statistically significant differences in the ideal and the actual entry points, inclined angles, and tailed angles. No screw was misplaced outside the pedicle of the three-dimensional printed model, and no ionizing X-ray radiation was used during screw placement under navigation. It is easy and safe to place C1/C2 pedicle screws under MR surgical navigation. Mixed reality-based navigation is feasible within upper cervical spinal fractures with improved safety and accuracy of C1/C2 pedicle screw insertion.

Keywords: Mixed reality; atlanto-axial; navigation; pedicle screw; three-dimensional print.

MeSH terms

  • Adult
  • Feasibility Studies
  • Female
  • Humans
  • Imaging, Three-Dimensional*
  • Intraoperative Period
  • Male
  • Middle Aged
  • Pedicle Screws*
  • Surgery, Computer-Assisted / instrumentation*
  • Young Adult