Study objective: When using a standard geometry laryngoscope, experts recommend engaging the hyoepiglottic ligament-a ligament deep to the vallecula not visible to the intubator. The median glossoepiglottic fold (hereafter termed midline vallecular fold) is a superficial mucosal structure, visible to the intubator, that lies in the midline of the vallecula. We aimed to determine whether engaging the midline vallecular fold with a standard geometry blade tip during orotracheal intubation improved laryngeal visualization.
Methods: We reviewed laryngoscopic videos from intubations by emergency physicians using standard geometry video laryngoscopes over a 2-year period. Two reviewers watched each video and recorded whether the blade tip engaged the midline vallecular fold (obscured the fold with the blade tip) and the best modified Cormack-Lehane grade and percent of glottic opening obtained. We compared laryngeal views in the presence and absence of fold engagement.
Results: We analyzed 183 discrete laryngoscopic episodes, including 113 instances in which the midline vallecular fold was engaged and 70 instances in which the fold was not engaged. The proportion with a Cormack-Lehane grade 1 or 2a was higher with fold engagement (96%) than without (87%) (absolute difference 9% [95% confidence interval (CI) 1 to 18%]). Ordinal logistic regression demonstrated that midline vallecular fold engagement was associated with a more favorable Cormack-Lehane grade (odds ratio 2.1 [95% CI 1.1 to 4.2]). The median percent of glottic opening score was 95% (interquartile range 90 to 100%) with fold engagement and 95% (65 to 100%) without engagement (median difference 0% [95% CI 0 to 5%]).
Conclusion: Engaging the midline vallecular fold with the laryngoscope blade tip during orotracheal intubation when using a standard geometry blade was associated with improved laryngeal visualization.
Copyright © 2021 American College of Emergency Physicians. Published by Elsevier Inc. All rights reserved.