Prostate MRI using a rigid two-channel phased-array endorectal coil: comparison with phased array coil acquisition at 3 T

Cancer Imaging. 2022 Mar 16;22(1):15. doi: 10.1186/s40644-022-00453-7.

Abstract

Background: To compare image quality, lesion detection and patient comfort of 3T prostate MRI using a combined rigid two-channel phased-array endorectal coil and an external phased-array coil (ERC-PAC) compared to external PAC acquisition in the same patients.

Methods: Thirty three men (mean age 65.3y) with suspected (n = 15) or biopsy-proven prostate cancer (PCa, n = 18) were prospectively enrolled in this exploratory study. 3T prostate MRI including T2-weighted imaging (T2WI) and diffusion-weighted imaging (DWI) was performed using an ERC-PAC versus PAC alone, in random order. Image quality, lesion detection and characterization (biparametric PI-RADSv2.1) were evaluated by 2 independent observers. Estimated signal-to-noise ratio (eSNR) was measured in identified lesions and the peripheral zone (PZ). Patient comfort was assessed using a questionnaire. Data were compared between sequences and acquisitions. Inter/intra-observer agreement for PI-RADS scores was evaluated.

Results: Twenty four prostate lesions (22 PCa) were identified in 20/33 men. Superior image quality was found for ERC-PAC compared to PAC for T2WI for one observer (Obs.1, p < 0.03) and high b-value DWI for both observers (p < 0.05). The sensitivity of PI-RADS for lesion detection for ERC-PAC and PAC acquisitions was 79.2 and 75% for Obs.1, and 79.1 and 66.7%, for Obs.2, without significant difference for each observer (McNemar p-values ≥0.08). Inter-/intra-observer agreement for PI-RADS scores was moderate-to-substantial (kappa = 0.52-0.84). Higher eSNR was observed for lesions and PZ for T2WI and PZ for DWI using ERC-PAC (p < 0.013). Most patients (21/33) reported discomfort at ERC insertion.

Conclusion: Despite improved image quality and eSNR using the rigid ERC-PAC combination, no significant improvement in lesion detection was observed, therefore not supporting the routine use of ERC for prostate MRI.

Keywords: Endorectal coil; Image quality; Magnetic resonance imaging; Prostate cancer; Signal to noise ratio (SNR).

MeSH terms

  • Aged
  • Diffusion Magnetic Resonance Imaging / methods
  • Humans
  • Magnetic Resonance Imaging / methods
  • Male
  • Prostate* / diagnostic imaging
  • Prostate* / pathology
  • Prostatic Neoplasms* / diagnostic imaging
  • Prostatic Neoplasms* / pathology
  • Signal-To-Noise Ratio