Canon CP-TX1 camera - As a screening tool for amblyogenic risk factors

Indian J Ophthalmol. 2022 Apr;70(4):1313-1316. doi: 10.4103/ijo.IJO_2161_21.

Abstract

Purpose: To evaluate the Canon CP-TX1 camera as a screening tool for ARFs in a pediatric population and estimate the prevalence of ARFs.

Methods: In a pediatric outpatient space, largely in the immunization clinic, after obtaining parental consent, we encouraged children to be photographed from a distance of 5 feet in a dim room by using a CP-TX1 camera with the red-eye reduction feature off. Based on the captured red reflex, children were labeled as normal (symmetrical red reflexes in the two eyes, with no visible crescents); all others were considered as abnormal or positive for ARFs. All photographed children were assessed by an optometrist/refractionist for VA by age-appropriate methods. Data were entered into a 2 × 2 contingency table on statpages.org, and diagnostic indices were calculated with 95%CI.

Results: With a sample of 262 children, we obtained a sensitivity of 0.82, a specificity of 0.98, a positive predictive value of 0.92, a negative predictive value of 0.94, a positive likelihood ratio of 41.06, a negative likelihood ratio of 0.17, and a prevalence of 0.24 for ARFs.

Conclusion: CP-TX1 performed well as a screening tool to identify ARFs in children. Placing such a camera in an immunization clinic offers a chance to identify children with ARFs at a time when amblyopia is eminently reversible.

Keywords: Amblyogenic risk factors; Brückners reflex; Immunization clinic.

MeSH terms

  • Amblyopia* / diagnosis
  • Amblyopia* / epidemiology
  • Child
  • Humans
  • Predictive Value of Tests
  • Risk Factors
  • Sensitivity and Specificity
  • Vision Screening* / methods