Methods for the calibration of bone conduction transducers at frequencies from 5 to 20 kHz

J Acoust Soc Am. 2022 May;151(5):2945. doi: 10.1121/10.0010381.

Abstract

Techniques for standardizing the output of bone conduction transducers over the 5-20 kHz range are presented. The techniques include definitions of the standard artificial mastoid (AM) impedance and force sensitivity in that high-frequency range using an impedance head coupled to a vibration source. The AM impedance is shown to vary with the contact area of the vibration source. The AM force sensitivity does not vary with the contact area but does exhibit sharp frequency dependences over the 14-20 kHz range. The sharp frequency dependence complicates the use of the AM force sensor as a stand-alone calibration device at these high frequencies. An alternative calibration scheme that uses an accelerometer interposed between the vibrator and the AM impedance is described. Comparisons of the two schemes demonstrate that the accelerometer method produces more consistent results. Comparisons of the force and acceleration output of one bone conduction vibrator at high frequencies suggest those outputs depend on the driven load. The loads used in the two calibration schemes are compared to what is known of the impedance of the skin-covered head.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Bone Conduction*
  • Calibration
  • Mastoid
  • Transducers*
  • Vibration