Validation of Segmental Post-PCI Physiological Gradients With IVUS-Detected Focal Lesions and Stent Underexpansion

JACC Cardiovasc Interv. 2023 Jul 24;16(14):1763-1773. doi: 10.1016/j.jcin.2023.03.044. Epub 2023 Jun 21.

Abstract

Background: Segmental postpercutaneous coronary intervention (PCI) pressure gradients may detect residual disease and potential targets for optimization. However, universal definitions of relevant segmental gradients are lacking.

Objectives: This study sought to evaluate the diagnostic performance of post-PCI fractional flow reserve (FFR), distal coronary pressure-to-aortic pressure ratio (Pd/Pa), and diastolic pressure ratio (dPR) gradients to detect residual focal lesions and stent underexpansion as observed by intravascular ultrasound (IVUS).

Methods: Patients from the IVUS-guided optimization arm of the FFR REACT (FFR-guided PCI Optimization Directed by High-Definition IVUS Versus Standard of Care) trial with complete IVUS and FFR pullback data were included. Patients with angiographically successful PCI and post-PCI FFR <0.90 underwent FFR, Pd/Pa, and IVUS pullbacks. dPR was calculated offline using dedicated software. Segmental pressure gradients (distal, in stent, and proximal) in segments ≥5 mm were evaluated against IVUS-detected residual disease (distal or proximal focal lesions and stent underexpansion).

Results: A total of 139 vessels were included (mean post-PCI FFR: 0.83 ± 0.05, range 0.56-0.89). Focal distal and proximal lesions were detected by IVUS in 23 (17.4%) of 132 and 14 (12.6%) of 111 vessels, respectively, whereas stent underexpansion was present in 86 (61.9%) vessels. Diagnostic ability of segmental FFR gradients to predict IVUS-detected distal and proximal lesions was moderate-to-good (area under the curve [AUC]: 0.69 and 0.84, respectively) and poor to moderate for segmental Pd/Pa and dPR gradients (AUC ranging from 0.58 to 0.69). In-stent gradients had no discriminative ability to detect stent underexpansion (FFR AUC: 0.52; Pd/Pa AUC: 0.54; dPR AUC: 0.55).

Conclusions: In patients with post-PCI FFR <0.90, segmental post-PCI pressure gradients have moderate discriminative ability to identify IVUS-detected focal lesions but no discriminative ability to identify IVUS-detected stent underexpansion.

Keywords: fractional flow reserve; intravascular ultrasound; percutaneous coronary intervention.

MeSH terms

  • Coronary Angiography
  • Coronary Artery Disease* / diagnostic imaging
  • Coronary Artery Disease* / therapy
  • Coronary Vessels / diagnostic imaging
  • Fractional Flow Reserve, Myocardial*
  • Humans
  • Percutaneous Coronary Intervention* / adverse effects
  • Predictive Value of Tests
  • Stents
  • Treatment Outcome
  • Ultrasonography, Interventional

Associated data

  • NTR/NL6523