Datawarehouse-enabled quality control of atrial fibrillation detection in the stroke unit setting

Heliyon. 2023 Jul 19;9(8):e18432. doi: 10.1016/j.heliyon.2023.e18432. eCollection 2023 Aug.

Abstract

Objective: (1) To assess the accuracy of a standard operating procedure (SOP) regarding the utilization of atrial fibrillation (AF) alarms in everyday clinical practice, and (2) to evaluate the performance of automated continuous surveillance for atrial fibrillation (AF) in hospitalized acute stroke patients.

Design: Retrospective cohort study.

Setting: Two stroke units from two tertiary care hospitals in Berlin, Germany.

Participants: We identified 635 patients with ischemic stroke diagnosis for the time period between 01. January and 30. September 2021 of which 176 patients had recorded AF alarms during monitoring. Of those, 115 patients were randomly selected for evaluation. After excluding 6 patients with hemorrhagic stroke in their records, 109 patients (mean age: 79.1 years, median NIHSS at admission: 6, 57% female) remained for analysis.

Intervention: Using a clinical data warehouse for comprehensive data storage we retrospectively downloaded and visualized ECG data segments of 65 s duration around the automated AF alarms. We restricted the maximum number of ECG segments to ten per patient. Each ECG segment plot was uploaded into a REDCap database and categorized as either AF, non-AF or artifact by manual review. Atrial flutter was subsumed as AF. These classifications were then matched with 1) medical history and known diseases before stroke, 2) discharge diagnosis, and 3) recommended treatment plan in the medical history using electronic health records.

Main outcome measures: The primary outcome was the proportion of previously unknown AF diagnoses correctly identified by the monitoring system but missed by the clinical team during hospitalization. Secondary outcomes included the proportion of patients in whom a diagnosis of AF would likely have led to anticoagulant therapy. We also evaluated the accuracy of the automated detection system in terms of its positive predictive value (PPV).

Results: We evaluated a total of 717 ECG alarm segments from 109 patients. In 4 patients (3.7, 95% confidence interval [CI] 1.18-9.68%) physicians had missed AF despite at least one true positive alarm. All four patients did not receive long-term secondary prevention in form of anticoagulant therapy. 427 out of 717 alarms were rated true positives, resulting in a positive predictive value of 0.6 (CI 0.56-0.63) in this cohort.

Conclusion: By connecting a data warehouse, electronic health records and a REDCap survey tool, we introduce a path to assess the monitoring quality of AF in acute stroke patients. We find that implemented standards of procedure to detect AF during stroke unit care are effective but leave room for improvement. Such data warehouse-based concepts may help to adjust internal processes or identify targets of further investigations.

Keywords: Alarm fatigue; Atrial fibrillation; Automated detection; Clinical data warehouse; ECG monitoring; Quality control; Stroke unit.