A Simple and Reliable 2D-Shear Wave Elastography and UltraSound Coefficient Attenuation Parameter Technique in Chronic Liver Diseases

Dig Dis Sci. 2024 Apr 28. doi: 10.1007/s10620-024-08444-3. Online ahead of print.

Abstract

Background: The performance and reliability criteria for Aixplorer MACH30 (SS) in chronic liver diseases (CLD) have not been validated.

Aims: The objectives were to define the optimal procedure, the accuracy for fibrosis and steatosis diagnosis, and the reliability criteria using SS.

Methods: Patients had 2D-shear wave elastography (SWE) and ultraSound-guided controlled attenuation parameter (SCAP) performed in triplicate at the mid-axillary line (MAL), posterior axillary line (PAL), and anterior axillary line (AAL). Performances of SWE and SCAP were defined using transient elastography (TE ≥ 9.5 kPa) and CAP (≥ 275 dB/m) using Fibroscan (FS) as reference and validated with liver biopsy (LB).

Results: FS and SS data from 203 CLD patients were analyzed (55 ± 14 years; 59% male; MASLD 58%). Median TE and CAP were 6.4 kPa (2.5-66.9) and 270 dB/m (141-400). The best technique for the diagnosis of advanced fibrosis and significant steatosis was the median of three SWE values and three SCAP values at MAL, PAL, and AAL with an AUROC of 0.96 [95% CI 0.93-0.98] and 0.91 [95% CI 0.86-0.95]. Only skin-to-liver distance ≥ 2.4 cm (p = 0.012, 95% CI 1.37-13.38) was independently associated with discordance. The accuracy of SWE (≥ 8.5 kPa) and SCAP (≥ 0.44) was analyzed in 58 patients with LB. The PPV and NPV were 50% and 94%, and 71% and 88% for fibrosis and steatosis, respectively.

Conclusion: A reliable diagnosis of advanced fibrosis and significant steatosis can be obtained with the median of three measurements in different liver portions using SS. The only non-reliable criterion is skin-to-liver distance ≥ 2.4 cm.

Keywords: Attenuation; Fibrosis; Non-invasive tests; Shear wave elastography; Steatosis; Transient elastography.