Ultrasound imaging guided precision histotripsy: Effects of pulse settings on ablation properties in rat brain

J Acoust Soc Am. 2024 Apr 1;155(4):2860-2874. doi: 10.1121/10.0025832.

Abstract

A high-frequency 6 MHz miniature handheld histotripsy device with an endoscopic form factor and co-registered high-resolution ultrasound imaging was developed. This device could allow precision histotripsy ablation during minimally invasive brain tumor surgeries with real-time image guidance. This study characterized the outcome of acute histotripsy in the normal in vivo rat brain using the device with a range of histotripsy pulse settings, including number of cycles, pulse repetition frequency, and pressure, as well as other experimental factors. The stability and shape of the bubble cloud were measured during ablations, as well as the post-histotripsy ablation shape in ultrasound B-mode and histology. The results were compared between histological images and the ultrasound imaging data to determine how well ultrasound data reflected observable damage in histology. The results indicated that while pulse settings can have some influence on ablation shape, sample-to-sample variation had a larger influence on ablation shape. This suggests that real-time ablation monitoring is essential for accurate knowledge of outcomes. Ultrasound imaging provided an accurate real-time indication of ablation shape both during ablation and post-ablation.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain* / diagnostic imaging
  • Brain* / pathology
  • Brain* / surgery
  • Equipment Design
  • High-Intensity Focused Ultrasound Ablation* / instrumentation
  • High-Intensity Focused Ultrasound Ablation* / methods
  • Male
  • Rats
  • Rats, Sprague-Dawley
  • Ultrasonography / methods
  • Ultrasonography, Interventional / methods