Space-Specific Mixing Excitation for High-SNR Spatial Encoding in Magnetic Particle Imaging

IEEE Trans Biomed Eng. 2024 May 13:PP. doi: 10.1109/TBME.2024.3400274. Online ahead of print.

Abstract

Objective: Magnetic Particle Imaging (MPI) is a radiation-free tracer-based imaging technology that visualizes the spatial distribution of superparamagnetic iron oxide nanoparticles. Conventional spatial encoding methods in MPI rely on a gradient magnetic field with a constant gradient strength to generate a field-free point or line for spatial scanning. However, increasing the gradient strength can enhance theoretical spatial resolution but also lead to a decrease in the Signal-to-Noise Ratio (SNR) and sensitivity of the imaging system. This poses a technical challenge in balancing spatial resolution and sensitivity, necessitating intricate hardware design.

Methods: To address this, we present a Space-Specific Mixing Excitation (SSME) technique for achieving high-SNR spatial encoding in MPI. By utilizing a dual-frequency excitation magnetic field with a non-homogeneous field strength, magnetic particles at each position generate unique intermodulation responses. By performing multi-channel acquisitions across the entire field of view, high SNR MPI signals can be acquired. When combined with reconstruction techniques based on system matrix, multi-dimensional SSME-MPI can be achieved.

Results: The effectiveness of the proposed method was validated through phantom and in vivo imaging experiments. The results demonstrate significant improvements in sensitivity (3.6-fold improvement) and spatial resolution (better than 1 mm) without any hardware modifications.

Conclusion: These findings demonstrate the capability of SSME to enhance both the spatial resolution and sensitivity of MPI.

Significance: This method provides a solution to the ongoing challenge of balancing spatial resolution and sensitivity in MPI, potentially facilitating the implementation of MPI in a wider range of medical applications.