Evaluation of human hemispheric blood flow based on non invasive carotid blood flow measurements using the range-gated Doppler technique

Stroke. 1982 May-Jun;13(3):392-8. doi: 10.1161/01.str.13.3.392.

Abstract

Common carotid blood flow (CCBF) was measured in 11 anesthetised patients without extracranial arterial disease (nine acute subarachnoid haemorrhages and two cases of head injury). The range-gated Doppler flowmeter with an adjustable range-gated time system and a double transducer probe was used to determine diameter, blood velocity, and blood flow of the common carotid artery. Values were, respectively, 5.9 +/- 1.1 mm, 13.8 +/- 6.1 cm . sec-1 and 387 +/- 183 cm3 . min-1. Mean cerebral blood flow (rCBF) measurements were simultaneously made by 133Xenon intra-arterial method, with a value of 22.28 +/- 5.96 ml . min-1 . 100 g. High correlation coefficient was found between CCBF and mean rCBF (r = 0.73, p less than 0.001). The range-gated Doppler flowmeter thus provides a non invasive and easily duplicated method for monitoring cerebral blood flow in anesthetized patients.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Blood Flow Velocity*
  • Brain Injuries / diagnosis
  • Carotid Arteries*
  • Cerebrovascular Circulation*
  • Female
  • Humans
  • Intracranial Aneurysm / diagnosis
  • Male
  • Middle Aged
  • Subarachnoid Hemorrhage / diagnosis
  • Ultrasonography*
  • Xenon Radioisotopes

Substances

  • Xenon Radioisotopes