Determination of free haemoglobin in serum by an automated assay using 4-aminophenazone and the Cobas Bio System

J Clin Chem Clin Biochem. 1981 Sep;19(9):971-6. doi: 10.1515/cclm.1981.19.9.971.

Abstract

A colorimetric method for the determination of free haemoglobin in human serum is described, using 4-aminophenazone as chromogen, and commercially available haemiglobin cyanide solutions as standards. The reaction system is suitable for manual measurements as well as determinations with a centrifugal analyser (Cobas Bio System). Accuracy, linearity and stability of the procedure is shown. A reference interval in sera of normal adults was established, with an upper limit of 5 mg/dl. The alternative use of heparin plasma instead of serum is discussed.

MeSH terms

  • Aminopyrine
  • Autoanalysis
  • Hemoglobins / analysis*
  • Humans
  • Kinetics
  • Spectrophotometry / methods

Substances

  • Hemoglobins
  • Aminopyrine