Limitations and late complications of third-generation automatic cardioverter-defibrillators

Circulation. 1995 Apr 15;91(8):2204-13. doi: 10.1161/01.cir.91.8.2204.

Abstract

Background: This study examines the limitations and complex management problems associated with the use of tiered-therapy, implantable cardioverter-defibrillators (ICDs).

Methods and results: The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57 +/- 11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients.

Conclusions: Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.

MeSH terms

  • Algorithms
  • Anti-Arrhythmia Agents / therapeutic use
  • Atrial Fibrillation
  • Death, Sudden, Cardiac / epidemiology
  • Defibrillators, Implantable* / adverse effects
  • Electrocardiography
  • Electrodes, Implanted / adverse effects
  • Equipment Design
  • Equipment Failure
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Survival Analysis
  • Tachycardia, Sinus
  • Tachycardia, Ventricular / epidemiology
  • Tachycardia, Ventricular / therapy*
  • Time Factors
  • Ventricular Fibrillation / epidemiology
  • Ventricular Fibrillation / therapy*

Substances

  • Anti-Arrhythmia Agents