Amputation of the fine guidewire tip during atherectomy using the Simpson "over-the-wire" peripheral atherectomy catheter

Cardiovasc Intervent Radiol. 1993 May-Jun;16(3):193-4. doi: 10.1007/BF02641892.

Abstract

After successful treatment of a localized eccentric atheromatous deposit in the distal superficial femoral artery using the Simpson "over-the-wire" atherectomy catheter, a short segment of guidewire tip was noted in the proximal peroneal artery. A digital angiogram of the distal vessels showed satisfactory run-off. There were no immediate clinical sequelae and none after 2 months follow-up.

Publication types

  • Case Reports
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Atherectomy / adverse effects*
  • Atherectomy / instrumentation
  • Equipment Failure
  • Femoral Artery
  • Humans
  • Male
  • Middle Aged